Site Preparation
Electrical Requirements
Table 2-4
Power consumption of 4072/4073 system instruments
Component
Flat-panel display

Main power cable

You must prepare the power cables to connect the 4072/4073 tester to the switchboard because it is not bundled
with the tester. This chapter describes how to prepare the main power cables.
Figure 2-11
Connecting power cable
Switchboard
Power Cable
When you connect the ground wire to the PDU of the system cabinet, use the terminal as shown in Figure 2-11.
Table 2-5 shows the main power cable specification. The cable must comply with the local regulation.
If you need to perform an ELECTRICAL SAFETY LEVEL CHECKS before the installation, see Appendix A ,
"Confirming the System Power Safety" to connect the site power line to the 4072/73 tester and check it.
Table 2-5
Specification for main power cable
Connection
Power line
type
option
AC 200/208 V
Flexible
AC 220/240 V
AC 200/208 V
Rigid/flexible
conduit
AC 220/240 V
NOTE
The torque specifications of the power cable connections at the PDU are 4.0 N-m for 208/220/240
V and 2.5 N-m for 200 V option.
26
AC 200 V
0.2 A
System Cabinet
5.5 mm
9.5 mm
Terminal
(E JST 5.5-5)
Power
Cable type
cable
SO
AWG 10
AWG 10
TEW
AWG 10
AWG 10
Agilent 4072/4073 Preinstallation Guide, Edition 4
Line voltage
AC 208 V
AC 220 V
0.2 A
0.2 A
400 mm
Ground
flexible cable
Neutral or line
Line
9 mm
Number of
lines
3
Use furnished cable clamp.
Opening has a diameter of
3
16.5 to 18.5 mm
Remove furnished cable
3
clamp. Opening has a
3
diameter of 19 mm.
AC 240 V
0.2 A
Cable clamp
Chapter 2