Table 2-3
Number of installable unit
Mounted system controller
c8000
(or C3750)
None
Table 2-4
Power consumption of 4072/4073 system instruments
Component
System cabinet
Testhead power supply (with testhead)
E4980A or 4284A presicion LCR meter
3458A digital multimeter
System controller (HP workstation c8000)
System controller (HP Visualize C3750)
DDS drive
81110A pulse generator (1 ea)
8114A pulse generator (1 ea)
E4411B spectrum analyzer
Color display (19 inch)
Chapter 2
Agilent 4072/4073 Preinstallation Guide, Edition 4
DDS drive rackmount
None
Yes
None
AC 200 V
1.0 A
2.5 A
0.8 A
0.4 A
3.9 A
3.5 A
0.6 A
1.5 A
2.5 A
1.5 A
1.4 A
Site Preparation
Electrical Requirements
81110A
8114A
E4411B
0
0
5
0
4
1
2
2
2
0
1
1
0
2
4
0
3
1
1
2
1
0
0
1
5
0
4
1
2
2
Line voltage
AC 208 V
AC 220 V
1.0 A
0.9 A
2.4 A
2.3 A
0.8 A
0.7 A
0.4 A
0.4 A
3.8 A
3.6 A
3.4 A
3.2 A
0.6 A
0.6 A
1.5 A
1.4 A
2.4 A
2.3 A
1.5A
1.4 A
1.3 A
1.2 A
0
0
0
0
1
1
1
0
0
0
1
1
1
1
1
AC 240 V
0.9 A
2.1 A
0.7 A
0.4 A
3.3 A
3.0 A
0.5 A
1.3 A
2.1 A
1.3 A
1.2 A