Confirming the System Power Safety
Connecting Site Power Line
5. Remove the disconnect device cover.
GND wire
GND terminal
7. Connect the ground wire to the ground terminal.
74
disconnect
cable clamp
device cover
6. Loosen (for flexible cable) or remove (for
rigid/flexible conduit) the cable clamp and pass the
main power cable through the cable clamp
opening.
power line wires
8. Connect line and neutral (or line) wires to the
terminal of the disconnect device. Use pozidrive
screwdriver (for 208/220/240 V) or
hex L screwdriver (for 200 V) to fasten the
terminal screws. The torque specifications of the
power cable connections at the PDU are 4.0 N-m
for 208/220/240 V and 2.5 N-m for 200 V option.
Agilent 4072/4073 Preinstallation Guide, Edition 4
Appendix A