Agilent Technologies 4072A Advanced Pre-Installation Manual page 33

Parametric
Hide thumbs Also See for 4072A Advanced:
Table of Contents

Advertisement

Figure 2-15 shows the dimensions of the bottom of the system cabinet. The figure shows the form and the position
of casters and leveling feet.
Figure 2-15
Dimensions of system cabinet bottom
Leveling foot
Caster
When the testhead is placed beside the manual wafer prober, strap the testhead to a table as shown in Figure 2-16.
Use suitable belts to secure the testhead according to the shape of the table.
Figure 2-16
Securing testhead
Manual
Wafer Prober
After an earthquake, confirm the system operation as follows:
1. Check the system visually for damage or disconnected parts.
2. Turn on the system and confirm that the tester starts correctly.
3. Execute the diagnostic program (diag4070) and check the measurement capability of the tester.
Chapter 2
Agilent 4072/4073 Preinstallation Guide, Edition 4
Leveling foot (4 place)
Caster
Caster
Testhead
Belt
Site Preparation
Safety Precautions
LEFT
BOTTOM
Table
33