Performance Tests
Transmission Test Set
STS-3/STM-1e O/P
136
155 Mb/s Output Mask Tests
1 Press <Menu>, choose Tx/Rx > Stored Settings then press
<Select>. Select Default then Recall.
2 Press <Menu>, choose Tx/Rx > Transmitter Settings then
press <Select>. Select the Physical tab. Set the Signal Rate
to STS-3.
3 Connect the 155 Mb/s OUT port to the Oscilloscope CHAN 1,
using the E2622A Termination Adapter Probe.
STS-3/STM-1e Output
BNC to BNC cable
54810A Oscilloscope
Installation and Verification Manual
CH1
E2622A
75/50 Ω
Adapter