140 Mb/S Output Mask Tests - Agilent Technologies J2126A Installation And Verification Manual

Transmission test sets
Hide thumbs Also See for J2126A:
Table of Contents

Advertisement

Transmission Test Set
140 Mb/s O/P
Installation and Verification Manual

140 Mb/s Output Mask Tests

1 Press <Menu>, choose Tx/Rx > Stored Settings then press
<Select>. Select Default then Recall.
2 Press <Menu>, choose Tx/Rx > Transmitter Settings then
press <Select>. Select the Physical tab. Set the Signal Rate
to 140 Mb/s.
3 Connect the 140 Mb/s OUT port to the Oscilloscope CHAN 1,
using the E2622A Termination Adapter Probe.
140 Mb/s UnBal Output
BNC to BNC cable
54810A Oscilloscope
Performance Tests
CH1
E2622A
75/50 Ω
Adapter
129
3