3
Performance Tests
104
Check that the optical power measurement reads between:
-31 dBm and -29 dBm
Press <Smart Test>, choose Shortcuts > Frequency
Measurement then press <Select>. The Transmission Test
Set now displays the received frequency results.
Check that the frequency measurement reads between:
51.83999 and 51.84001 MHz for [STM-0 OPT] or [OC-1]
155.51999 and 155.52001 MHz for [STM-1 OPT] or [OC-3]
Installation and Verification Manual