Performance Tests
Transmission Test Set
DS1 O/P
Bantam to Bantam Cable
116
DS1 Output Mask Test
1 Press <Menu>, choose Tx/Rx > Stored Settings then press
<Select>. Select Default then Recall.
2 Press <Menu>, choose Tx/Rx > Transmitter Settings then
press <Select>. Select the Physical tab. Set the Signal Rate
to DS1 as shown in the example below.
3 Connect the Transmission Test Set's DS1 OUT port to the
Oscilloscope using the E2621A Termination Adapter Probe,
set probe switch to 100 Ω.
54810A Oscilloscope
DS1 Balanced O/P
CH1
Set
E2621A
to
Bal/Unbal
100 Ω
Adapter
Installation and Verification Manual