2 Mb/s Unbalanced Output Mask Tests

1 Press <Menu>, choose Tx/Rx > Stored Settings then press
<Select>. Select Default then Recall.
2 Press <Menu>, choose Tx/Rx > Transmitter Settings then
press <Select>. Select the Physical tab. Set the Signal Rate
to 2 Mb/s, Termination 75 ohms (Unbal).
3 Connect the 2 Mb/s BAL OUT port to the Oscilloscope
CHAN 1, using the E2622A Termination Adapter Probe.
2 Mb/s UnBal Output
BNC to BNC cable
54810A Oscilloscope
Performance Tests
CH1
E2622A
75/50 Ω
Adapter
121