3
Performance Tests

Performance Test Record

NO TE
page 118page 118page 118118page 122
Page No.
Test Description
Performance Self Test
72
step 3
72
step 6
72
step 7
Internal Reference Clock Accuracy
75
step 4
77
step 13
78
step 2
External Clock Reference Inputs and Clock Reference Output
81
step 6
81
step 7
82
step 12
140
Model: Agilent J2126/7A Transmission Test Set
Location:
Tested by:
Temperature:
Humidity:
The test steps applicable are dependent on the instrument options
installed. Enter N/A if a test does not apply for your instrument.
Confidence Test
Frequency Measurement
Ethernet Test
(10/100/1G)
2 MHz Clock OUT
DS1 Clock OUT
Alternative DS1 Test
(Counter indication for
framed signal)
2 MHz Clock Ext
Signal Locked
2 Mb/s Data Ext Unbal
Serial No.:
Options:
Certified by:
Date:
Result
Min
Actual
Pass/Fail
Pass/Fail
Pass/Fail
2.0479908 MHz
–4.5 ppm
1.540993 MHz
2.047999999 MHz
Pass/Fail
2.047999999 MHz
Installation and Verification Manual
Max
2.0480092 MHz
+4.5 ppm
1.541007 MHz
2.048000001 MHz
2.048000001 MHz