Invacare® Robin®
1. Place the hoist on a surface (such as the corner of a
table) so that the hoist is supported yet the straps and
hand control are free hanging. Ensure the hoist is safe
and cannot slip/fall of the table.
2. Remove the two yellow cables A from the twin speed
switch B.
3. Remove the two screws C.
4. Attach the new twin speed switch with screws C.
Connect the yellow cables A. Ensure, they are
connected to the bottom and top terminals.
5. Re-assemble the hoist and reset the emergency stop.
20
3.12 Replace Contact Disc for Robin® Mover
CAUTION!
Static electricity can damage the PCBs inside the
hoist.
Before working in the inside of the hoist:
– Pull the red emergency strap to turn off the
hoist.
– Remove the hoist from the track system. (Refer
to 3.2 Dismounting the ceiling hoist, page 10)
– Remove the cover from the hoist. (Refer to 3.4
Removing the hoist cover, page 13)
– An anti-static mat must be used when working
inside the ceiling hoist.
Screwdriver
1586747-C