Invacare® Robin®
1. Remove the battery. Refer to steps 1 and 2 in chapter
3.6 Replacing the battery, page 14.
2. Remove all plugs B, C, D, E and F from the PCB A.
3. Release the PCB A from the retaining holders by gently
pressing down the locking clips and levering the PCB
slightly away from the chassis plate.
4. Lift the PCB up to remove.
5. Place the new PCB into the chassis and press it slightly
against the chassis plate until the retaining holders snap
into place.
6. Connect the plugs B, C, D, E and F to the new PCB.
7. Re-insert battery, re-assemble the hoist and reset the
emergency stop.
16

3.8 Replacing the fuse

CAUTION!
Static electricity can damage the PCBs inside the
hoist.
Before working in the inside of the hoist:
– Pull the red emergency strap to turn off the
hoist.
– Remove the hoist from the track system. (Refer
to 3.2 Dismounting the ceiling hoist, page 10)
– Remove the cover from the hoist. (Refer to 3.4
Removing the hoist cover, page 13)
– An anti-static mat must be used when working
inside the ceiling hoist.
1. Remove the hand control. Refer to chapter 3.3 Replacing
the hand control, page 13
2. Remove the battery. Refer to steps 1 to 3 in chapter 3.6
Replacing the battery, page 14.
3. Gently pull the fuse A straight out away from the
holder B.
1586747-C