Electrical System; Environmental Conditions - Invacare Robin Series Service Manual

Ceiling hoist
Hide thumbs Also See for Robin Series:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3

6.3 Electrical System

Maximum current
input
Protection class
1
(ceiling hoist: IP24; hand control: IP44)
Insulation class
Applied part complying with the
specified requirements for protection
against electrical shock according
Sound level
Number of lifts
65 lifts of 0.5 m with 80 kg
per charge
Intermittent
(periodic motor
10%, max. 2 min. / 18 min.
operation)
Battery
NiMH Battery 24 V / 1.9 Ah
Charging time
Lifting speed with
2.1 cm/s
200 kg load
Traverse speed in
the tracks
1586747-C
Robin®
Robin® Mover
0.9 A
Entire device: IPx4
Class II equipment
Type B applied part
to IEC60601-1.
50 – 55 dB (A)
Up to 3 hours at 20°C
2.1 cm/s
n/a
14 cm/s
Charger
During charging: 25 – 36 V DC
Voltage output
100 – 240 V AC, 50 – 60 Hz
Voltage supply
Protection class
1
IP41 or IP67 (depending on version)
1
See product label and label on each electric device for
correct protection class. The lowest IP classification decides
the overall classification of the device.
IPx4: Protected against water splashed from any
direction.
IP24: Protected against objects larger than 12.5 mm and
protected against water splashed from any direction.
IP44: Protected against objects larger than 1 mm and
protected against water splashed from any direction.
IP41: Protected against objects larger than 1 mm and
protected against vertically dripping water.
IP67: Dust-tight and protected against temporary
immersion into water.

6.4 Environmental conditions

Storage and
transportation
Temperature
-30 °C to +50 °C
Relative humidity
10 % to 70 %
Atmospheric
pressure
Technical Data
No-load: 41 V DC
Operation
+5 °C to +40 °C
20 % to 90 % at 30
°C, not condensing
700 hPa to 1060 hPa
33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Robin moverRobin

Table of Contents