Technical Specifications - 3M Versaflo TR-800 Series User Instructions

Powered air purifying respirator (papr) belt mounted assemblies
Hide thumbs Also See for Versaflo TR-800 Series:
Table of Contents

Advertisement

Available languages

Available languages

3M™ Battery Pack TR-830 can provide approximately 750 full charge/discharge cycle equivalents while maintaining 80% of
its original capacity, when used under recommended conditions and during its first year of service. Normal use of the battery
and natural aging of battery cells will gradually decrease the battery's available capacity. For additional information on
battery use and optimization see the Versaflo™ TR-830 Chargers and Battery Packs User Instructions and 3M Technical Data
Bulletin #223 3M™ Versaflo™ PAPR Battery Maintenance.
Do not store battery packs where temperatures may exceed 122°F (50°C).
Disposal
Dispose of lithium-ion battery packs according to local environmental regulations. Do not dispose of in standard waste
bins, in a fire or send for incineration. Failure to properly dispose of battery packs may lead to environmental
contamination, fire or explosion and may result in serious injury or death.
In the United States and Canada, 3M participates in the Call2Recycle program which provides for 3M™ PAPR Batteries to be
dropped off (free of charge) at participating recycling collection sites. Please dispose of batteries according to federal, state,
provincial and local regulations. Contact the Call2Recycle Hotline at 1-877-273-2925 or go to their website,
www.call2recycle.org, for drop off site locations.
Used filter/cartridges should be disposed of based on the contaminants collected on it and according to local environmental
regulations.
The motor/blower and charger contain electronic components. Dispose of according to local and government regulations.

TECHNICAL SPECIFICATIONS

Consult the 3M™ Versaflo™ Powered Air Purifying Respirator TR-800 Technical Specifications sheet for additional
specifications.
Airflow
Intrinsic Safety
Weight:
TR-800 motor/blower- no battery pack
Standard battery pack- TR-830
High durability belt- TR-626
Easy clean belt- TR-627
Operating temperature
Operating altitude/elevation range
Storage conditions:
Relative humidity
Temperature (daily storage)
Temperature (extended storage)
Optimal temp for battery packs to maintain existing
charge while stored off charger
Battery charging*
* Based on the internal temperature of the battery
Shelf life from date of manufacture when stored in original
packaging and recommended storage conditions:
1. Motor/blower
2. Battery pack
3. Filters and cartridges
W WARNING
Approx. 6.7 cfm (190 lpm): Standard flow
Approx. 7.2 cfm (205 lpm): Medium flow
Approx. 7.8 cfm (220 lpm): High flow
Exia Division 1: IS Class I, II, III; Division 1
(Includes Division 2); Groups C, D, E, F, G; T4
Ex ia I Ma
Class I, Zone 0, AEx ia IIB; T4
Class I, Zone 0, Ex ia IIB; T4
Zone 20, AEx ia IIIC; T135°C
Zone 20, Ex ia IIIC; T135°C
-20°C ≤ T
≤ 55°C
a
Approx. 1.5 lbs (675 grams)
Approx. 1.5 lbs (663 grams)
Approx. 0.9 lbs (425 grams)
Approx. 0.6 lbs (270 grams)
23°F to 131°F (-5°C to 55°C)
-328 feet to 16404 feet (-100 meters to 5000 meters)
< 90%
-22°F to 122°F (-30°C to 50°C)
40°F to 95°F (4°C to 35°C)
59°F (15°C)
Range: 32°F to 104°F (0°C to 40°C)
Optimal: 68°F to 77°F (20°C to 25°C)
1. 5 years*
2. 6 months
3. 5 years
*Run for 5 minutes/year