Download Print this page

Cleaning And Storage; Specifications; Listing Of Components, Accessories And Replacement Parts - 3M Adflo 35-5401-01 Manual

Powered air purifying respirator (papr) assembly

Advertisement

c. The low air flow or low battery alarms trigger:
• The air flow alarm activates when air flow falls below 6 CFM (170 lpm) for greater than 15-30 seconds. The
3M™ Adflo™ PAPR will shut down 2 minutes after continuous alarm activation.
• The battery alarm - during approximately the last 15 to 30 minutes of charge, the bottom bar on
motor/blower battery LED will flash continuously one second ON, one second OFF. The audible alarm will
beep four times in two seconds, repeating every 30 seconds. The final 15 to 60 seconds the alarm will fast
beep every 1/2 second. 3M™ Adflo™ PAPR will shut down 30 seconds after the quick beep starts.
d. In the event only an audible or only a visual alarm triggers, the user should still immediately leave the
contaminated area.
e. Breathing becomes difficult.
f.
You feel dizzy or your vision is impaired.
g. You taste or smell contaminants.
h. Your face, eyes, nose or mouth become(s) irritated.
i.
You suspect that the concentration of contaminants may have reached levels at which this respirator may no
longer provide adequate protection.
5. Do not remove the respirator or reach your hand into the headgear in areas where the air is contaminated.
6. Follow exiting and decontamination procedures specific to your workplace for turning off the motor/blower and
removing the respirator system.
The 3M™ Adflo™ PAPR is not intrinsically safe. Do not use in flammable or explosive atmospheres, doing so may
result in serious injury or death.
Do not enter a contaminated area until properly donning the respirator system. Do not remove or turn off the respirator
before leaving the contaminated area. Doing so may result in sickness or death.
Do not use the 3M™ Adflo™ PAPR system with a heavily loaded spark arrestor or without the spark arrestor screen
properly installed. Exposing a heavily loaded spark arrestor, or filter assembly to direct contact with sparks or molten
metal spatter may damage the filter, allowing unfiltered air into the breathing zone, which may result in sickness or
death, and may cause the filter or blower housing to ignite, resulting in serious injury, sickness or death.

CLEANING AND STORAGE

Dispose of lithium ion battery packs according to local environmental regulations. Do not crush, disassemble, dispose
of in standard waste bins, in a fire or send for incineration. Failure to properly dispose of battery packs may lead to
environmental contamination, fire or explosion.
The 3M™ Adflo™ PAPR Assembly should be cleaned and inspected after each use and prior to storage.
NOTE: To avoid damage, do not use solvents for cleaning. Do not immerse in water or spray directly with water.
Cleaning
Detach the battery pack, breathing tube and headgear from the motor/blower. Inspect all parts for damage or other signs
of excessive wear. Replace all damaged parts prior to storage or next use.
1. Motor/blower and battery pack: Clean the outer surfaces of the 3M™ Adflo™ PAPR Assembly and battery pack with a soft
cloth dampened in a solution of water and mild, pH neutral detergent. Do not immerse the motor/blower or battery pack in
water. Do not use solvents or abrasive cleaners. Do not attempt to clean the interior of the motor/blower with compressed
air or vacuum. Dust accumulation in the motor/blower could indicate possible filter leakage or improper storage and must
be addressed prior to next use. Ensure the electrical contacts of the motor/blower and battery pack are dry.
2. Breathing tube: Clean the exterior and connection sites on the breathing tube with the water and detergent solution.
Optional breathing tube covers can also be used to facilitate cleaning. Ensure the breathing tube is completely dry
before using or storing. The L-122SG and SG-50W breathing tube interiors contain sound absorbing material. They
cannot be immersed in liquids for cleaning and must be replaced if wet. Wiping the exterior is insufficient.
3. HE filter: Open the filter cover and inspect the HE filter, pre-filter and spark arrestor. Replace if excessively dirty, wet or
damaged. The HE and pre-filters cannot be cleaned. The spark arrestor can be cleaned using a clean, soft cloth dipped
in a solution of water and a mild pH neutral detergent. Completely dry the spark arrestor with a clean cloth. If the spark
arrestor cannot be cleaned or is damaged, replace with a new spark arrestor.
Inspection
Follow inspection procedures listed in this User Instructions.
Motor/Blower Storage
The 3M™ Adflo™ PAPR system should be stored at a temperature between -4°F to 131°F (-20°C to 55°C). The motor/
blower should be stored in a clean environment out of direct sunlight.
Battery Pack Storage
To help maximize battery pack service life:
• Store the battery pack at -4°F (-20°C) to 115°F (45°C). R.H. < 85%. Storage of the battery pack outside of this range
will shorten service life of the battery.
• The 3M™ Adflo™ PAPR Battery Pack may remain on the charger for 3 to 4 days after completion of charging, but for
maximum battery life, disconnect the charger after a full charge has been received.
• Battery packs for long-term storage should be disconnected from the motor/blower.
• Battery packs not intended for immediate or regular use should be charged upon receipt and placed on regular
recharge schedule (every 3 months of non-use). Prolonged storage of battery packs without regular recharge may
damage the battery pack cells.

SPECIFICATIONS

Technical
Air flow
Weight (approximate)
1. Assembly (motor/blower; HE filter; belt;
battery pack and SG-30W breathing tube)
2. OV/AG Cartridge
3. Nuisance Odor Pad Assembly
Motor/Blower Assembly
1. Operating Temperature
2. Storage Temperature (RH <90%)
3. Operating Altitude and Limitation
4. Noise level
Belt Size (Approximate)
Belt Extender
WARNING
WARNING
Greater than 6 CFM (170 lpm)
1. 3.6 lbs (1.65 kg)
2. 1.72 lbs. (0.78 kg)
3. 0.320 lbs (0.145kg)
1. 23ºF to 131ºF (-5º to 55ºC)
2. -4°F to 131°F (–20°C to 55°C)
If 3M™ Adflo™ PAPR system is stored at temperatures below
32°F (0°C), the battery pack must be allowed to warm up to
achieve optimal battery performance.
3. Altitude*:
• Less than 2500 feet (800 meters) MSL –HE filter and
OV/AG Cartridge, HE filter and nuisance odor pad assembly,
or HE filter alone may be used.
• 2500 to 3500 feet (800 to 1200 meters) MSL – HE filter
alone may be used.
• Greater than 3500 feet (1200 meters) MSL – not
recommended. Use the high altitude version of the 3M™
Adflo™ PAPR (35-3099-01) or a different 3M™ PAPR.
* If you have questions regarding these products contact 3M
Technical Service.
4. Noise levels – less than 80 dBA at user's ear excluding external
noise.
29.5 to 45.25 inches (75 to 115 cm)
Approximately 26 inches (66 cm)
Battery Pack/Charger
1. Chemistry
2. Run Time
3. Charging Temperatures
4. Storage Temperatures
5. Battery Pack Service Life
6. Charger Codes
Motor/Blower Alarms
Low air flow
Low battery pack voltage
Hazardous Location (Intrinsic Safety)
Shelf Life – HE Filter, Nuisance Odor Pad
Assembly and OV/AG Cartridge
Assigned Protection Factor
Refer to the User Instructions for the specific head gear to be used to determine the assigned protection factor (APF) for
the 3M™ Adflo™ PAPR system. Consult 3M™ Technical Data Bulletin #175 for additional information on APFs and
supporting test data.
Service Life
Chemical Cartridges
The useful service life of the 3M™ Adflo™ PAPR Organic Vapor/Acid Gas Cartridge, 15-0399-99 or 15-0499-99, will
depend on the specific type, volatility and concentration of the contaminants and environmental conditions such as
humidity and temperature. Replace cartridges in accordance with an established change schedule, regulations, or
sooner if smell, taste, or irritation from contaminants are detected.
Nuisance Odor Pad Assembly
The useful service life of the 3M™ Adflo™ PAPR Nuisance Odor Pad will depend on the environmental conditions and
concentration of the contaminants and environmental conditions such as humidity and temperature. The nuisance odor
pad should only be used for concentrations of organic vapors not exceeding the OSHA permissible exposure limit (PEL) or
applicable government occupational exposure limits, whichever is lower. Replace if smell or taste is detected.
Particle Filter
The 3M™ Adflo™ PAPR Motor/Blower Assembly is equipped with a five-step particle filter-loading indicator. There are three
green and two red indicators. The illumination of each indicator light signifies increased filter clogging and a reduction in
battery run time. When the red indicator(s) are illuminated, battery run time will be significantly reduced and HE particle filter
and pre-filter replacement is recommended. When using a chemical cartridge or nuisance odor pad assembly with a fresh
particle filter, the first two LEDs of the scale will be illuminated due to the additional resistance to air flow.
The HE filter should also be changed if it becomes wet or damaged. The pre-filter should be changed at least as frequently
as the HE filter. The pre-filter and HE filter should be changed if the battery pack operating time becomes shortened or if
the low air flow alarm is activated.

LISTING OF COMPONENTS, ACCESSORIES AND REPLACEMENT PARTS

Part Number
Description
Battery Packs and Chargers
35-1099-07
Battery Pack, Li
35-0099-08
Lithium Ion Battery Smart Charger
Belts
15-0099-16
Belt, Leather
15-0099-06
Belt Tongue Extender Replacement, Leather
35-0099-14
Belt, Nylon webbing with quick connection
TR-325
Belt, Versaflo, Standard
TR-326
Belt, Versaflo, High Durability
BPK-01
Backpack
Breathing Tubes
15-0099-10
Breathing tube - 9000 HWR, 9000 Flexview and ClearVisor only
L-122SG
Breathing tube - L-SG series headgear only
SG-30W
Breathing tube - Self adjusting - 9100 series and M-100/300 headgear
SG-40W
Breathing tube - Heavy duty - 9100 series and M-100/300-series headgear
SG-50W
Breathing tube – Sound dampening – 9100 series helmets. Required when used with
3M™ Adflo™ PAPR with HE Filter and OV/AG Cartridge and Welding Helmet 9100-Air,
9100 MP, and must be used with M-400 headgear when used with any 3M™ Adflo™
PAPR configuration.
Filters and Cartridges – Refer to the NIOSH approval label for proper filter/cartridge/headgear combinations
15-0099-99X06
Pre-filter, 6-pack
15-0099-99X12
Pre-filter, 12-pack
15-0299-99X02
High Efficiency Particulate Filter, 2/case
15-0299-99X06
High Efficiency Particulate Filter, 6/ case
15-0299-99X36
High Efficiency Particulate Filter, 36/case
1. Lithium Ion (Li)
2. •
10-12 hours*
• 5.5 hours with OV/AG cartridge*
*New battery pack and HE filter, battery life will be reduced if using
at altitudes greater than 3500 feet.
3. 32°F to 104°F (0°C to 40°C). If the charger's LED blinks rapidly
during charging, the battery pack is either too hot or too cold to
charge.
4. -4°F (-20°C) to 115°F (45°C). R.H. < 85%. Storage of the
battery pack outside of this range will shorten service life of the
battery.
5. Approximately 500 charges.
6. Charger Code Table:
Mode
Description
LED Indication
Yellow Permanent on
Yellow
Standby
Green
Yellow flashing slow
Pre-Charge
Yellow
Green
Yellow
Waiting
Y
ellow and g
reen
f
or correct
alterna
ting
Green
tempera
ture
Rapid charge
Green flashing fast
Yellow
Green
Maintenance
Green flashing slow
Yellow
charge
Green
Ready
Green permanent on
Yellow
Green
Error
Yellow flashing fast
Yellow
Green
Activates when air flow falls below 6 CFM (170 lpm) for greater
than 15-30 seconds. 3M™ Adflo™ PAPR will shut down 2 mins.
after continuous alarm activation.
During approximately the last 15 to 30 minutes of charge, the
bottom bar on motor/blower battery LED will flash continuously one
second ON, one second OFF. The audible alarm will beep four times
in two seconds, repeating every 30 seconds. The final 15 to 60
seconds the alarm will fast beep every 1/2 second. 3M™ Adflo™
PAPR will shut down 30 seconds after quick beep starts.
The 3M™ Adflo™ PAPR Assembly is not approved for use in
hazardous or intrinsically safe locations.
5 years from date of manufacture.

Advertisement

loading

This manual is also suitable for:

Adflo 34-5401-02sgAdflo 35-3099-01