Updating Software; Batteries - Invacare Storm4 Series Service Manual

Hide thumbs Also See for Storm4 Series:
Table of Contents

Advertisement

Invacare® Storm⁴ Series
Installing Light Board
1. Install parts in reverse order.
2. Plug all plugs into their old positions.
3. Check all mobility device functions.
6.9.7 Drive Program Selection After Component Replacement
The drive program is saved in the remote, and also in the power module. If one of these two components is replaced,
the system must be told which of the components has not been replaced so that it knows which contains the current
controller profile.
The system allows normal usage of the mobility device after component replacement only if the profile is selected
afterwards.
For details, see corresponding remote manual.

6.9.8 Updating software

Invacare is continuously carrying out further development and improvements to drive programs for powered mobility
devices. For this reason, you should always check whether the drive program version number is up-to-date when carrying
out any repairs or regular maintenance.
If a new version is available, the drive program must be updated. The procedure for updating the drive program is
described in the wizard software operating manual.
The power wheelchair is supplied with a standard drive program. If the drive program has been customised,
you have to perform this customisation again, after installing the new drive program. This also applies to the
customer-specific options of the seat setting for ACS2 remotes, which are activated ex works.
WARNING!
Any changes to the drive program can affect the driving characteristics and the tipping stability of the
mobility device.
– Changes to the drive program may only be carried out by trained Invacare specialist dealers.
– Invacare can only give a warranty for safe mobility device driving behaviour - especially the tipping stability -
for unaltered standard drive programs.
Dynamic Wizard software
Manual for Wizard software
Further requirements, such as a minimum system configuration for the PC used for programming, required
programming cables etc. can be taken from the Wizard software operating manual.
When an electrical adjustment option is retrofitted, such as electrical legrests, then this option needs to be activated
in the drive program as well if you have an ACS2 remote. For more information, refer to the user manual of the
Wizard software and the installation instructions for the electronic modules.

6.10 Batteries

CAUTION!
Injury hazard and possible material damages if batteries are handled improperly
The installation of new batteries may only be carried out by authorised specialists.
– Observe the warning information on the batteries.
– Only use battery versions stated in the specifications.
CAUTION!
Fire and burns hazard if battery terminal is bypassed
– Take great care to ensure that the battery terminals are never short-circuited with tools or mechanical
mobility device parts.
– Ensure that the battery terminal caps have been replaced if you are not working on the battery terminals.
44
1555830-C

Advertisement

Table of Contents
loading

Table of Contents