CAUTION!
Risk of fire and burns
Fitting incorrect strip fuse causes fire hazard.
– Only fix strip fuses in sequence shown in image below.
– Tighten nuts with 3.3 or 3.5 Nm.
1
Strip fuse
2
Ring terminal
3
DIN 6923 nut
•
8 mm socket wrench
•
Strip fuse
•
Flat screwdriver
•
Torque wench 0–20 Nm (or similar)
If the fuseholder is damaged, you can replace it completely with the battery cables.
Removing Main Fuse
1. Turn off mobility device.
2. Remove battery shroud, disconnect battery plugs and
pull batteries out of battery box. See Making Batteries
Accessible.
3. Fuse holder (1) is located on top of batteries.
4. Pry open snap hook B with flat screwdriver C. Fuse
holder shroud A is open.
Older fuse holder versions may be locked with tie
wrap. If so, cut open tie wrap to access fuse strip.
1555830-C
Fig. 6-39 Correct sequence of fitting original strip fuse
Fig. 6-40
Fig. 6-41
Service