Invacare Storm4 Series Service Manual page 47

Hide thumbs Also See for Storm4 Series:
Table of Contents

Advertisement

CAUTION!
Risk of fire and burns
Fitting incorrect strip fuse causes fire hazard.
– Only fix strip fuses in sequence shown in image below.
– Tighten nuts with 3.3 or 3.5 Nm.
1
Strip fuse
2
Ring terminal
3
DIN 6923 nut
8 mm socket wrench
Strip fuse
Flat screwdriver
Torque wench 0–20 Nm (or similar)
If the fuseholder is damaged, you can replace it completely with the battery cables.
Removing Main Fuse
1. Turn off mobility device.
2. Remove battery shroud, disconnect battery plugs and
pull batteries out of battery box. See Making Batteries
Accessible.
3. Fuse holder (1) is located on top of batteries.
4. Pry open snap hook B with flat screwdriver C. Fuse
holder shroud A is open.
Older fuse holder versions may be locked with tie
wrap. If so, cut open tie wrap to access fuse strip.
1555830-C
Fig. 6-39 Correct sequence of fitting original strip fuse
Fig. 6-40
Fig. 6-41
Service
47

Advertisement

Table of Contents
loading

Table of Contents