AGFA DX-G User Manual page 90

Hide thumbs Also See for DX-G:
Table of Contents

Advertisement

DX-G/DX-M
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
Tests of Resistance
to Disruption
Conducted high fre-
quency disturbance
variables in accor-
dance with
IEC 61000-4-6
Radiated high fre-
quency disturbance
variables in accor-
dance with
IEC 61000-4-3
90
IEC 60601
Level of
Test Level
Agreement
3 V
3 V
eff
eff
150 kHz
to 80 MHz
3 V/m
3 V/m
80 MHz
to 2.5 GHz
Electromagnetic Environment
Use portable and mobile radio sets at
a safe distance from the device
(including the leads) not closer than
the recommended protective dis-
tance, which is calculated according
to the equation suitable for the trans-
mission frequency.
Recommended protective distance:
d = 1.2
P
d = 1.2
80 MHz to 800 MHz
P
d = 2.3
800 MHz to 2.5 GHz
P
With P as the rated power of the
transmitter in watts (W) in accor-
dance with the manufacturer infor-
mation on the transmitter and d as
the recommended protective dis-
tance in metres (m).
The field strength of stationary radio
transmitters is lower than the level of
a
the agreement
at all frequencies in
accordance with an on-site investiga-
b
tion
.
Disruptions are possible near devices
that carry the following symbol:
2321 E EN 20130722

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dx-m

Table of Contents