Power consumption (typical values)
220 V - 240 V / 50-60 Hz configuration
Standby
Average continuous scanning
During scanning
100 V - 120 V / 50-60 Hz configuration
Standby
Average continuous scanning
During scanning
Environmental conditions (during operation)
In line with IEC721-3-2: 2M2 and 2K2 with following restrictions:
Room temperature
Maximum temperature change rate
Relative humidity
Light
Environmental conditions (during storage)
Room temperature
Relative humidity
Barometric pressure
Environmental conditions (during transport)
Room temperature
Relative humidity
Barometric pressure
Environmental conditions for mobile installation (during transport)
In line with IEC721-3-5: 5M1 and 5K1 with following restrictions:
Room temperature
2321 E EN 20130722
87 W
237 W
590 W max.
92 W
245 W
621 W max.
Agfa CR HM5.x Mammo: between
+20 °C and +30 °C (68 - 86°F)
Other plates and cassettes: between
+15 °C and +30 °C (59 - 86°F)
max. 0.5 °C/min (0.9°F/min)
between 15 % and 75 %
(non-condensing)
no exposure to direct sunlight, max. 2500 Lux
between -25 °C and 55 °C
between 10 % and 95 %
between 70 kPa and 106 kPa
between -25 °C and 55 °C
maximum 85 %
minimum 70 kPa
between 10 °C and 40 °C
DX-G/DX-M
79