DX-G/DX-M
5
Switch on the digitizer.
Image quality check after transport
The image quality check must be performed after installation of the digitizer in
a mobile environment and is recommended to be repeated after transport.
The check is done with a flat field exposure and should be performed with a
cassette of the largest format used at customer site.
X-Ray source
General radiography
24
Note: If the digitizer is switched on without removing the transport lock on
the front side, the following warning message is displayed: "Transport locks
are attached. Switch off the digitizer and remove transport locks."
Exposure conditions
It is recommended to expose the cassette with 2 expo-
sures of 10 µGy or 1 mR each. Rotate the cassette 180°
after the first exposure to compensate for the heel effect.
Typical settings for 10 µGy or 1 mR are:
75 kV
12 mAs
130 cm SID
large focus
1.5 mm Copper filter
Identify the cassette as System Diagnosis GenRad - Flat
Field.
Table 3: Image quality check after transport