D Series Syringe Pumps
Section 8 Serial Interface
8-18
this example, the master and the network controller are a
single unit. If they were separate units, the master would
send the inquiry to the network controller; then the
network controller would send the message to the slave
unit the next time it is polled. The slave would respond
with the message to the network controller. The next time
the master is polled, the message would be relayed. The
same sequence would occur with all messages. Since the
master and the network controller are combined in this
example, the relaying of messages is not necessary.
Unit 6
R027SERIES=1240-02__, Model ___D PUMP, REV __XX[CR]
The pump responds with identity and software revision
letter. (In this example 02__ would be 021; Model ___D
would be 260D; REV __ signifies the software revision, XX
would be replaced by the correct checksum, which is B4.)
Network Controller and Master Unit #1
6R006REMOTE16[CR]
This places the pump in the remote mode.
Unit 6
R 8E[CR]
The pump acknowledges that it accepted the command.
Network Controller and Master Unit #1
6R00ACONST FLOWF8[CR]
This puts the pump into constant flow rate mode.
Unit 6
R 8E[CR]
The pump verifies that it received the message.
Network Controller and Master Unit #1
6R009FLOW=1.00AB[CR]
This sets the pump's flow rate to 1.00 ml per minute.
Unit 6
R 8E[CR]
The pump verifies that it received the message.
Network Controller and Master Unit #1
6R 58[CR]
Polls the pump.
Unit 1
R 8E[CR]
Pump responds.
Network Controller and Master Unit #1
6R003RUNF0[CR]
The pump is started.
Unit 1
R 8E[CR]
The pump responds.