Appendix B Continuous Flow Introduction, Installation, and Operation for 30D
B.3.2 Dual Electric Valve
Installation
Plumbing connections
Electrical connections
To install the electric valve package (refer to Figure B-2 on the
following page):
1. Position the valve bases 1.3 cm apart.
2. Use the plugs to stopper the ports that will not be con-
nected.
RISK OF INJURY. THE PRESSURE PRODUCED
COULD BE 2068 BAR. PLEASE UTILIZE
APPROPRIATE TUBING AND CONNECTIONS
NOTED IN THE MANUAL.
3. Loosely attach the tubing lengths from the valve assembly
to the pumps.
4. Install the four bracket panhead screws on the bottom of
the bracket.
5. Tighten the tubing nuts.
6. Connect the inlet tube to the supply reservoir. Connect the
outlet tee to your apparatus.
7. Connect the DB-25 cable to the controller rear panel "elec-
tric valves" connector (Figure B-2).
D Series Syringe Pumps
DANGER
B-5