Two Controllers (Compiler Required); Three Or More Controllers (Compiler Required) - Teledyne D series Installation And Operation Manual

Table of Contents

Advertisement

D Series Syringe Pumps
Section 8 Serial Interface
Figure 8-1 Serial network connection example - Single connection
8.4.2 Two Controllers
(Compiler required)
Figure 8-2 Serial network connection example - Dual connection
8.4.3 Three or More
Controllers (Compiler
required)
8-4
Null Modem Cable
part #480-7996-00
With this configuration, your sample program is capable of:
• Constant flow
• Constant pressure
• Gradient
To connect two controllers in the network, use the 'Y' cable
#68-1020-198, as shown in Figure 8-2.
Note
In order for the network to operate properly, all instruments
connected to the network must be turned on even if they are
not being used.
If additional instruments are to be connected in series, use the
daisy chain cable, #68-1020-180. This cable is attached to the
remaining connector on the 'Y' cable (Figure 8-2), and then to the
RS-232-C connectors on the rear panels of the additional con-
trollers.