Installation
2.4.2 Temperature Control
Jacket
Figure 2-12 System with temperature control jacket installed
Section 2 Fluid System Connections and Accessories
Notches in the cover provide openings for the inlet and outlet
tubing, and for the cable to the pressure transducer.
1. Install the back cover.
2. Route the cable around tubing through the appropriate
tabs.
3. Install the front cover and secure the side latches.
The optional cylinder temperature jacket assists in maintaining
cylinder temperatures (–30 to 100°C) by circulating liquids, such
as water or water/ethylene glycol solution, through the
and lower hose connectors.
CO
The jacket is very useful for SFC applications where cylinder
cooling facilitates filling with fluids such as liquid CO
jacket can also be used with a circulating temperature-controlled
bath to keep the fluid inside the pump at a constant temperature.
This may be necessary when operating at very low flow rates
(below 500 µl/min), where temperature fluctuation can cause
flow variations.