Appendix B Continuous Flow Introduction, Installation, and Operation for 30D
B.5.4 No Check Valve
Figure B-7 Keystrokes to set up No Check Valve
To limit fluid backflow from the pump to the supply reservoir one
of two approaches is generally used. Either a check valve is
installed in the supply line or the N
software feature is enabled on the controller.
The N
C
V
O
HECK
ALVE
inlet valves such that, just prior to refill, the pump is depres-
surized to a user-specified value (pressure point) or until a
user-specified time out period (time point). This reduces pressure
in the cylinder before the inlet valve is opened.
The NCV software feature is disabled by default; however, if no
check valve is installed on the supply line, it is recommended
that this feature is enabled.
1. Contin Const Flow
2. Contin Const Press
(O
(
)
MORE
(
-
)
MULTI
PUMP
1. Pressure Point
1. AB Pumps
2. Time Point
2. CD Pumps
3. Function NCV
D Series Syringe Pumps
C
O
HECK
(NCV) feature controls the opening of the
O
)
THER
PTIONS
(NCV)
V
(NCV)
ALVE
B-11