2.2.4 Accessory Packages

Installation
Section 2 Fluid System Connections and Accessories
Wear goggles to perform the following procedure.
Tools required: Fine jewelers file, goggles, and two pairs of pliers
1. Using the jewelers file, score the tubing around its entire
circumference.
2. Secure the tubing with pliers on either side of the score
line with approximately 1.5mm between each pair of pliers
and the score line. Do not squeeze the tubing too tightly, as
this will flatten or deform its exterior.
3. Bend the tubing back and forth to crack it at the score line.
4. It may be necessary to deburr the outer tubing ends with
the file. Make sure the tubing ends are clean and the inner
bore is clear before installing the cut tube.
Note
It is often impossible to remove a burr that blocks the inner
bore.
The accessory packages for all pumps contain fittings that allow
you to attach tubing to the pump (refer to Figure 2-1), as well as
appropriately sized cylinder seals. Note that the tubing is not
included.
1. Slide first the nut and then the ferrule over the tubing.
2. Push the tubing all the way into the port.
3. Hold the tubing in place and tighten the nut.
Table 2-1 D Series Syringe
Pump Accessory Packages
Pump Model
1000D
500D/HPx
65D
30D
260D
100DM/DX
65DM
D Series Syringe Pumps
CAUTION
Part Number
60-1249-103
60-1249-016
60-1249-024
60-1249-131
60-1249-012
60-1249-015
60-1249-028
2-3