Figure 3-15 Output range selection jumpers

3.14.2 Current Loop Output

Section 3 Basic Programming and Operation
5V and the preferred output scale is 1 volt per 10 ml/min,
is set at 50 ml/min.
FLOW
specification.
A controller running one to two pumps can also output cylinder
volume on the remaining outputs. The volume output range is
not adjustable. Full scale is equal to one pump stroke.
Table 3-2 Analog Output Options (connections to
female 25 pin Sub-D)
Pin No.
Description
1
Chassis (earth)
2
Analog common
3
Analog common
4
Analog common
5
Analog common
14
Output #1
15
Output #2
16
Output #3
17
Output #4
For additional information about analog output options,
download technical bulletin TB34Syringe Pump Analog Output
Options.
The pump controller can be configured to provide a 4–20 mA
current loop output. The 4-20mA analog output board is
pre-installed at the factory, but can also be purchased in an
upgrade kit. The board initially contains three output cards, for
flow rate, pressure, and volume for Pump A. Additional cards can
D Series Syringe Pumps
cannot exceed the pump
MAX FLOW
One or Two
Three Pump
Pump Function
Function
Flow Rate A
Flow Rate A
Flow Rate B
Flow Rate B
Volume A
Flow Rate C
Volume B
Volume B
MAX
3-23