Table 7-5 Coils (Continued)
Register
Address
00038
37
00039
38
00040
39
00041
40
00042
41
00043
42
00044
43
00045
44
00046
45
00047
46
00048
47
00049
48
00050
49
00051
50
00052
51
00053
52
00054
53
00055
54
00056
55
00057
56
00058
57
00059
58
00060
59
00061
60
00062
61
00063
62
00064
63
00065
64
00066
65
00067
66
00068
67
00069
68
00070
69
00071
70
00072
71
00073
72
00074
73
00075
74
00076
75
00077
76
Type
Size
Description
R
1 BIT
1: Pump C is Equilibrating
R
1 BIT
1: Pump D is Equilibrating
R
1 BIT
1: Pump A is in Hold (clock stopped)
R
1 BIT
1: Pump B is in Hold (clock stopped)
R
1 BIT
1: Pump C is in Hold (clock stopped)
R
1 BIT
1: Pump D is in Hold (clock stopped)
R
1 BIT
1: Pump A & Pump B in Concentration Gradient
R
1 BIT
1: Pump A in Pressure Gradient
R
1 BIT
1: Pump A in Flow Gradient
R/W
1 BIT
1: Controller in Local Operation
R/W
1 BIT
1: Controller in Remote Operation
R/W
1 BIT
1: Controller in External Operation
R/W
1 BIT
1: External Refill control enabled
R/W
1 BIT
1: Zero pressure sensor offset for Pump A
R/W
1 BIT
1: Zero pressure sensor offset for Pump B
R/W
1 BIT
1: Zero pressure sensor offset for Pump C
R/W
1 BIT
1: Zero pressure sensor offset for Pump D
R/W
1 BIT
0: Pressure Integrator for Pump A OFF 1: Pressure Integrator for Pump A ON
R/W
1 BIT
0: Pressure Integrator for Pump B OFF 1: Pressure Integrator for Pump B ON
R/W
1 BIT
0: Pressure Integrator for Pump C OFF 1: Pressure Integrator for Pump C ON
R/W
1 BIT
0: Pressure Integrator for Pump D OFF 1: Pressure Integrator for Pump D ON
R/W
1 BIT
0: Pressure Deadband for Pump A OFF 1: Pressure Deadband for Pump A ON
R/W
1 BIT
0: Pressure Deadband for Pump B OFF 1: Pressure Deadband for Pump B ON
R/W
1 BIT
0: Pressure Deadband for Pump C OFF 1: Pressure Deadband for Pump C ON
R/W
1 BIT
0: Pressure Deadband for Pump D OFF 1: Pressure Deadband for Pump D ON
R/W
1 BIT
0: AUTO REFILL A OFF 1: AUTO REFILL A ON
R/W
1 BIT
0: AUTO REFILL B OFF 1: AUTO REFILL B ON
R/W
1 BIT
0: AUTO REFILL C OFF 1: AUTO REFILL C ON
R/W
1 BIT
0: AUTO REFILL D OFF 1: AUTO REFILL D ON
R/W
1 BIT
0: AUTO FILL A OFF 1: AUTO FILL A ON
R/W
1 BIT
0: AUTO FILL B OFF 1: AUTO FILL B ON
R/W
1 BIT
0: AUTO FILL C OFF 1: AUTO FILL C ON
R/W
1 BIT
0: AUTO FILL D OFF 1: AUTO FILL D ON
R/W
1 BIT
0: DIGITAL OUTPUT BIT 1 HIGH 1: DIGITAL OUTPUT BIT 1 LOW
R/W
1 BIT
0: DIGITAL OUTPUT BIT 2 HIGH 1: DIGITAL OUTPUT BIT 2 LOW
R/W
1 BIT
0: DIGITAL OUTPUT BIT 3 HIGH 1: DIGITAL OUTPUT BIT 3 LOW
R/W
1 BIT
0: DIGITAL OUTPUT BIT 4 HIGH 1: DIGITAL OUTPUT BIT 4 LOW
R/W
1 BIT
0: DIGITAL OUTPUT BIT 5 HIGH 1: DIGITAL OUTPUT BIT 5 LOW
R/W
1 BIT
0: DIGITAL OUTPUT BIT 6 HIGH 1: DIGITAL OUTPUT BIT 6 LOW
R/W
1 BIT
0: DIGITAL OUTPUT BIT 7 HIGH 1: DIGITAL OUTPUT BIT 7 LOW
D Series Syringe Pumps
Section 7 Modbus Configuration
7-5