D Series Syringe Pumps
Section 7 Modbus Configuration
Read/
Register Address
Write
40009
8
R/W
40011
10
R/W
40013
12
R/W
40015
14
R/W
40017
16
R/W
40019
18
R/W
40021
20
R/W
40023
22
R/W
40025
24
R/W
40027
26
R/W
40029
28
R/W
40031
30
R/W
40033
32
R/W
40035
34
R/W
40037
36
R/W
40039
38
R/W
40041
40
R/W
40043
42
R/W
40045
44
R/W
40047
46
R/W
40049
48
R/W
40051
50
R/W
40053
52
R/W
40055
54
R/W
40057
56
R/W
40059
58
R/W
40061
60
R/W
40063
62
R/W
40065
64
R/W
40067
66
R/W
40069
68
R/W
7-8
Table 7-6 Holding Registers (Continued)
# of
Type
Description
Registers
Pressure set-point for multi-pump pair AB in
FLOAT
2
Continuous Flow Constant Pressure mode.
Pressure set-point for multi-pump pair CD in
FLOAT
2
Continuous Flow Constant Pressure mode.
Flow rate set-point for Pump A in Constant
FLOAT
2
Flow mode.
Flow rate set-point for Pump B in Constant
FLOAT
2
Flow mode.
Flow rate set-point for Pump C in Constant
FLOAT
2
Flow mode.
Flow rate set-point for Pump D in Constant
FLOAT
2
Flow mode.
Flow rate set-point for multi-pump pair AB in
FLOAT
2
Continuous Flow Constant Flow mode.
Flow rate set-point for multi-pump pair CD in
FLOAT
2
Continuous Flow Constant Flow mode.
FLOAT
2
Refill flow rate for Pump A.
FLOAT
2
Refill flow rate for Pump B.
FLOAT
2
Refill flow rate for Pump C.
FLOAT
2
Refill flow rate for Pump D.
FLOAT
2
Maximum pressure set-point for Pump A.
FLOAT
2
Maximum pressure set-point for Pump B.
FLOAT
2
Maximum pressure set-point for Pump C.
FLOAT
2
Maximum pressure set-point for Pump D.
FLOAT
2
Minimum pressure set-point for Pump A.
FLOAT
2
Minimum pressure set-point for Pump B.
FLOAT
2
Minimum pressure set-point for Pump C.
FLOAT
2
Minimum pressure set-point for Pump D.
FLOAT
2
Maximum flow rate set-point for Pump A.
FLOAT
2
Maximum flow rate set-point for Pump B.
FLOAT
2
Maximum flow rate set-point for Pump C.
FLOAT
2
Maximum flow rate set-point for Pump D.
FLOAT
2
Minimum flow rate set-point for Pump A.
FLOAT
2
Minimum flow rate set-point for Pump B.
FLOAT
2
Minimum flow rate set-point for Pump C.
FLOAT
2
Minimum flow rate set-point for Pump D.
Maximum flow limit in Constant Pressure
FLOAT
2
mode for Pump A.
Maximum flow limit in Constant Pressure
FLOAT
2
mode for Pump B.
Maximum flow limit in Constant Pressure
FLOAT
2
mode for Pump C.
Units
ATM; BAR; kPa; PSI See NOTE 1.
ATM; BAR; kPa; PSI See NOTE 1.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ATM; BAR; kPa; PSI See NOTE 1.
ATM; BAR; kPa; PSI See NOTE 1.
ATM; BAR; kPa; PSI See NOTE 1.
ATM; BAR; kPa; PSI See NOTE 1.
ATM; BAR; kPa; PSI See NOTE 1.
ATM; BAR; kPa; PSI See NOTE 1.
ATM; BAR; kPa; PSI See NOTE 1.
ATM; BAR; kPa; PSI See NOTE 1.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.
ml/min; ml/hr; ul/min; ul/hr See NOTE 2.