Omron CS1W-AD041-V1 Operation Manual page 279

Sysmac cs/cj series
Table of Contents

Advertisement

Analog Input Functions and Operating Procedures
Input signal
to the Unit (V)
Conversion data
Response Time
Symbols
Calculation Example
256
the delay in the conversion data in comparison to changes in the input sig-
nals will be as shown in the following diagram.
2. Specify "no mean value processing" to follow conversion of a rapid change
in input signals.
Time (ms)
Time (ms)
t: Delay
Unit: ms
m
64
4
258.5
130.5
3
196.5
100.5
2
134.5
70.5
1
67
35
m: Number of input words used in DM Area
n: Average number of buffers set for the input number for which to find the
response time
The following example calculations are for a resolution of 8,000 with an appli-
cation using inputs 1 and 4, 64 averaging buffers set for input 1, and no aver-
aging set for input 4.
• Response time for input 1: t = {(64 − 2) x 2 + 10.5} = 134.5 (ms)
• Response time for input 1: t = 1 x (2 + 2) = 4 (ms)
For V = 20 V ( − 10 to 10 V)
Using One Word
t = n + (2 to 3)
Using m Words (1 ≤ m ≤ 8)
No averaging (n = 1) or two averaging buffers (n = 2):
t = n x (m + 2)
n averaging buffers (4 ≤ n ≤ 64):
t = (n − 2) x m + 10.5
n
32
16
8
66.5
34.5
52.5
28.5
38.5
22.5
19
11
Section 6-6
4
2
1
18.5
12
6
16.5
10
5
14.5
8
4
7
5