Viewing Test Limits; If The Test Limit Or Fiber Type Is Not Valid With The Test; Entering Job Settings - Fluke OF-500-01 OptiFiber Technical Reference Handbook

Of-500 series optifiber
Hide thumbs Also See for OF-500-01 OptiFiber:
Table of Contents

Advertisement

Viewing Test Limits

To see the values of a test limit that includes the fiber type
selected on the Cable tab, do the following:
S
1
Press
; then select a test's tab.
2
Select TEST LIMIT.
3
Highlight a limit on the SET TEST LIMIT screen; then
A
press
View Limit.
If the Test Limit or Fiber Type is Not Valid with the
Test
The fiber type was changed on the Cable tab in Setup,
and the current test limit does not include the fiber
type. Select a different test limit.
The fiber type or test limit is missing a wavelength the
test uses. For example, you are trying to run a dual-
wavelength OTDR test, but the fiber type or limit does
not include both wavelengths. Select a different fiber
type or test limit.
The installed module does not support the selected
fiber type or test limit. For example, a singlemode fiber
type is selected, but a multimode module is installed.
Select a different fiber type or test limit or install a
different module.
Getting Started Testing Fiber
The test limit does not include parameters for the type
of test you are trying to run. For example, you are
trying to run an OTDR test, but only loss/length limits
are defined in the selected test limit. Select a test limit
that includes limits for the test you want to run.
If you selected a custom fiber type or limit, you could also
edit the fiber type or limit to make it valid with the test. See
Chapter 10 for details.

Entering Job Settings

Job settings let you identify and organize the test results
you save. To access the job settings press
describes the job settings.
See "Tutorial: Preparing to Save Tests" on page 2-10 for
examples of how to configure the job settings.
General Test Settings
S
. Table 2-2
2
2-7

Advertisement

Table of Contents
loading

Table of Contents