Determining Test Limits For Other Calibration Intervals - Fluke 5700A Series Service Manual

Multi-function calibrator
Hide thumbs Also See for 5700A Series:
Table of Contents

Advertisement

5700A/5720A Series II Calibrator
Service Manual
Item
No.
9.
Current Shunt Adapter
10.
Frequency Counter
11.
Standard Resistors (1)
12.
DC Current Shunt (2)
13.
AC/DC Current Shunt
14.
Metal Film Resistors
15.
Differential Amplifier
16.
Distortion Analyzer
17.
Kelvin-Varley Voltage
Divider
18.
HF Spectrum Analyzer
(used in optional test
for wideband
distortion)
1: A DMM may be used for all but the 1
using an external current source is used for low-value resistance.
2: Needed only for 5725A Amplifier testing.

3-40. Determining Test Limits for Other Calibration Intervals

The verification procedures in this chapter test to the 90 day 99% confidence
specification limits. For other calibration intervals it is necessary to calculate new
specification limits and, if necessary, new test limits.
The following examples show how the 90 day limits were calculated. These examples
illustrate how you can calculate the specifications and test limits, if necessary, for other
calibration intervals.
3-48
Table 3-15. Minimum Use Requirements (cont)
Description
Used in conjunction with 5790A and A40-
series shunts to facilitate AC Current
measurements
10 Hz to 30 MHz ±0.002%
0.1Ω nominal, true value certified to within
20 ppm, rated for 2A DC; 1Ω nominal, true
value certified to within 6 ppm; 1.9Ω
nominal, true value certified to within 6 ppm;
10Ω nominal, true value certified to within 6
ppm; 1 kΩ nominal, true value certified to
within 5.5 ppm; 10 kΩ nominal, true value
certified to within 3.5 ppm; 19 kΩ nominal,
true value certified to within 4 ppm; 10 MΩ
nominal, true value certified to within 15
ppm; 19 MΩ nominal, true value certified to
within 28 ppm
Range: 10A
Uncertainty: ±0.008%
Ranges: 20 mA, 200 mA, 2A and 10A
Frequency: 10 Hz to 10 kHz
Uncertainty: ±310 ppm at 10 Hz; ±100 ppm
at 20 Hz; ±50 ppm at 40 Hz, 1 kHz; ±100
ppm at 5 kHz, 10 kHz
Values: 200Ω, 2 kΩ, and 1 MΩ Temperature
C°: T9 or better Power Rating: 1/4 Watt
Tolerance: ±1%
Sensitivity: 5 µV rms Bandwidth selectable
to 10 kHz
Range: 2V to 300V
Frequency: 10 Hz to 600 kHz
Ratio uncertainty: ±0.1 ppm of input
Freq. Range: 2 MHz to 120 MHz Input
Level: 3V (+20 dBm to -60 dBm)
Minimum Use Specifications
and 1.9
values. For those values using the DMM, a test method
Recommended
Equipment
Fluke 792A-7004
Philips PM 6669
Fluke 742A-1
Fluke 742A-1.9
Fluke 742A-10
Fluke 742A-1k
Fluke 742A-10k
Fluke 742A-19k
Fluke 742A-10M
Fluke 742A-19M
L&N 4221B (0.1Ω)
Fluke Y5020
Fluke A40-20 mA
Fluke A40-200 mA
Fluke A40-2A
Fluke A40A-10A (2)
Stock Items
Tektronix 7A22 w/7000-
Series Mainframe
Krohn-Hite 6900B
Fluke 720A
HP 8590A