Setting the Reference in Far End Source Mode

1
Turn on the tester and source and let them warm up
for 5 minutes.
2
On the Loss/Length tab in Setup set REMOTE END
SETUP to Far End Source.
3
Clean the connectors on the source and the reference
test cords.
F
4
Press
; then select Set Loss/Length Reference.
5
Make the connections shown on the SET REFERENCE
screen and in Figure 7-12.
6
Turn on the far end source and verify the output is set
to the correct wavelength and is in continuous-wave
mode.
7
Highlight a wavelength; then press
7-30
8
9
See "About Setting the Reference" on page 7-10 for more
information about referencing.
t
.
Optional: From the TEST SETUP screen, you can enter
reference test cord lengths to meet TSB-140 reporting
requirements:
L
Use
to highlight a reference test cord number,
t
press
; then enter a length. Press
done.
C
Press
OK to leave the TEST SETUP screen.
W
Caution
Do not disconnect the reference test cord from the
tester or source after setting the reference. If you
do, you must set the reference again to ensure
valid measurements.