Teledyne API - Model T200H/T200M Operation Manual
6.3.4. CHANGING THE NO
The converter operates at 315º C. Avoid severe burns by allowing the assembly to cool. Do
not handle the assembly until it is at room temperature. This may take several hours.
07270D DCN7141
CONVERTER
2
The NO
converter is located in the center of the instrument, see Figure 3-5 for location,
2
and Figure 6-3 for the assembly. The converter is designed for replacement of the
cartridge only; the heater with built-in thermocouple can be reused.
1. Turn off the analyzer power, remove the cover and allow the converter to cool.
2. Remove the top lid of the converter as well as the top layers of the insulation until
the converter cartridge can be seen.
CAUTION – HOT SURFACE!
3. Remove the tube fittings from the converter.
4. Disconnect the power and the thermocouple of the converter. Unscrew the
grounding clamp of the power leads with a Phillips-head screw driver.
5. Remove the converter assembly (cartridge and band heater) from the can. Make a
note of the orientation of the tubes relative to the heater cartridge.
6. Unscrew the band heater and loosen it, take out the old converter cartridge.
Figure 6-3:
7. Wrap the band heater around the new replacement cartridge and tighten the screws
using a high-temperature anti-seize agent such as copper paste. Make sure to use
proper alignment of the heater with respect to the converter tubes.
8. Replace the converter assembly, route the cables through the holes in the can and
reconnect them properly. Reconnect the grounding clamp around the heater leads
for safe operation.
NO
Converter Assembly
2
Instrument Maintenance
209