Das Structure; Table 4-7: Das Data Channel Properties - Teledyne T200H User Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API - Model T200H/T200M Operation Manual

4.7.1. DAS STRUCTURE

4.7.1.1. DAS Channels
PROPERTY
NAME
The name of the data channel.
TRIGGERING
The event that triggers the data channel to
EVENT
measure and store the datum
NUMBER AND
A User-configurable list of data types to be
LIST OF
recorded in any given channel.
PARAMETERS
The amount of time between each channel data
REPORT PERIOD
point.
The number of reports that will be stored in the
NUMBER OF
data file. Once the limit is exceeded, the oldest
RECORDS
data is over-written.
Enables the analyzer to automatically report
RS-232 REPORT
channel values to the RS-232 ports.
CHANNEL
Enables or disables the channel. Allows a channel
ENABLED
to be temporarily turned off without deleting it.
Disables sampling of data parameters while
CAL HOLD OFF
instrument is in calibration mode
1
More with APICOM, but only the first six are displayed on the front panel).
2
When enabled records are not recorded until the DAS HOLD OFF period is passed after calibration mode. DAS HOLD OFF set in
the VARS menu (see Section 4.12.)
07270D DCN7141
The DAS is designed around the feature of a "record", an automatically stored single
data point. (e.g. concentration, PMT signal level, etc.). Records are organized into data
channels which are defined by properties that characterize the:
Type of date recorded (e.g. concentration, PMT signal level, etc.);
Trigger event that causes the record to be made (e.g. every minute, upon exiting
calibration mode, etc.);
How many records to be stored, as well as;
How the information is to be stored (e.g. average over 1 hour, individual points,
minimum value over last 5 minutes, etc.).
The configuration of each DAS channel is stored in the analyzer's memory as a script,
which can be edited from the front panel or downloaded, edited and uploaded to the
instrument in form of a string of plain-text lines through the communication ports.
The key to the flexibility of the DAS is its ability to store a large number of
combinations of triggering events and data parameters in the form of data channels.
Users may create up to 20 data channels. For each channel one triggering event is
selected and one or all of the T200H/M's 25 data parameters are allowed. The number
of parameters and channels is limited by available memory.
The properties that define the structure of an DAS data channel are:

Table 4-7: DAS Data Channel Properties

DESCRIPTION
DEFAULT
"NONE"
ATIMER
1 - PMTDET
000:01:00
100
OFF
ON
OFF
2
.
Operating Instructions
SETTING RANGE
1.
Up to 6 letters or digits
Any available event
(see Appendix A-5).
Any available parameter
(see Appendix A-5).
000:00:01 to
366:23:59
(Days:Hours:Minutes)
1 to 1 million, limited by
available storage space.
OFF or ON
OFF or ON
OFF or ON
79