Power-On Self-Best (Post) - Nellcor N-395 Operator's Manual

Table of Contents

Advertisement

WAWING: Each time the monitor is used, check alarm
limits to ensure that they are appropriate for the patient
being monitored.
Important! Prior to using the N-395, carefully read this manual,
accessory directions for use, all precautionary information in
boldface type, and all specifications.
Before using the N-395 in a clinical setting, verify that the
monitor is workiilg properly and is safe to use. Proper working
condition can be verified by successful completion of the power-
on self-test described in the following steps, and by following
instructioi~s contained in the "Monitoring Mode" paragraph of
this section.
Ensure that the supply voltage selector switch on the rear panel
matches the AC voltage at your location.
Power-On Self-Test (POST)
WARNING: Ensure that the speaker is clear of any
obstruction. Failure to do so could result in an inaudible
alarm tone,
1. Plug an appropriate Nellcor sensor firmly into the SCP-10
or MC- 10 cable aizd lower the SCP- 10 or MC- 10 sensor
lock over the sensor connector until it clicks into place.
Insert the SCP-I0 or MC-I0 into the N-395 seiisor port.
Apply the sensor to the patient as described in the sensor
directions for use.
2. Turn on the N-395 by pressing the POWER ONiOFF
Button. The inonitor automatically starts a power-on self-
test (POST), which tests its circuitry.