Nellcor N-395 Operator's Manual page 42

Table of Contents

Advertisement

%SpO2
Seconds
Satseconds
2 x
2
=
4
4 x
3
=
12
6 x
6
=
36
Total SatSeconds
=
52
After approxi~nately 10.9 seconds the SatSeconds alarm would
sound, because 50 SatSeconds had been exceeded. See arrow
(+)
in Figure 5.
0
1
2 3 4 5 6 7 8 9 1 0 1 1
SECONDS
Figure
11:
Alarm Response with Satseconds
Saturation levels may fluctuate rather than remaining steady for
a period of several seconds. Often, %Sp02 levels may fluctuate
above and below the alarm limit, re-entering the non-alann
range several times.