Nellcor N-395 Operator's Manual page 96

Table of Contents

Advertisement

Data Porf Protocol
Data Source
N-395
VERSION
2.0.0.0
CRC:
XXXX
SpQ2
limit:
70-109%
PR
Limit:
60-160BPM
TIME
%Sp02
BPM
PA
Status
Data in the highlighted box above represents the model number
of the monitor, in this case the N-395.
Software Revisioil Level
N-395
VERSION 2.0.0.0
CRC: XXXX
Sp02 Limit: 7 0 * + i O 0 %
PR Limit: 60-260BPM
TIME
%Sp02
BPM
PA
Status
The next data field tells tlie user the software level, (Versioi~
2.0.0.0) and a software verification ilumber (CRC: XXXX).
Neither of these numbers should change during normal operation.
The numbers may change if the monitor is serviced and receives
a software upgrade.
Alarm Limits
-
,
- . .
N-395
VERSION 2.0.0.0
CRC:
XXXX
Sp02 Limit: 70-100%
PR Limit: 60-16OBPM
TIME
%SpQ2
BPW
PA
Status
The last data field in the top line indicates the high and the low
alarm limits for %Sp02 and for the pulse rate (PR). In the
example above the low alarm liinit for Sp02 is 70% and the high
alarm liinit is 100%. Pulse Rate alarm limits are 60 and 160
bpin.
Column Headings
N-395
VERSION
2.0.0.0
CRC:
XXXX
SpQ2 Limit:
70-900%
PR
L i m i t :
60-IBOBPM
TIME
%Sp02
BPM
PA
Status
Actual c01um1 headings are in the second row of the column
heading line. Patient data presented i 1 1 tlie chart, from left to
right, is the:
time the patient data was obtained
e
current %SpOz value
e
current Pulse Rate
curreilt Pulse Amplitude
operating status of the N-395.