WARNING: Do not silence the N-395 audible alarm or
decrease its volume if patient safety could be compromised.
The SpaceLabs monitor provides both audible and visual alarm
indications. Silencing the N-395 alann will also silence the
SpaceLabs monitor alarms. The monitors must be able to sound
an audible alarm in order to inaintain patient safety.
Marquette Communications
The N-395 sends Sp02, pulse rate, and alann status data to the
Marquette monitor.
The Marquette inonitor requires an Octanet
TM
interface module
to interface with the N-395 pulse oximeter. The interface
module comes with an interface cable, GE Marquette part
number 4 1796 1-033, that connects to the Nellcor interface
cable.
The RS-232 hardwire interface cable has a DB-15 connector
for the N-395 and the applicable connector for the Marquette
Octanet
TM
interface module cable. Nellcor cable part number
902254 is recoinmended for this interface.
Note:
Spare parts for the N-395 are listed on the Internet,
confirm the cable part nuinber before ordering a cable.
The Internet address is:
Corrupt data will be indicated by a Communications Error
displayed on the Marquette monitor. The Marquette monitor
will detect corrupt data in less than 7 seconds.
When the N-395 is in the Marquette inode of operation the
interface baud rate must be set to 9,600 bits per second.
Pressing the SETUP softkey, then the NEXT softkey, a ~ l d then
the COMM softltey to select BAUD. Use the ADJUST UP or
ADJUST DOWN buttons to select the correct baud rate.
The GE Marquette monitor only sounds audible alarms.
Silencing the N-395 audible alarm has no effect on the GE
Marquette monitor sounding an alarm.