Setting The Analog Display Range - Konica Minolta CA-310 Instruction Manual

Display color analyzer
Table of Contents

Advertisement

2. Setting the Analog Display Range

The analog display section displays the difference (%) between the measured value and the target color as well as
the difference (%) between measured values in the case of a measurement mode other than flicker mode**.In the
case of flicker mode, the measured values will be displayed as they are.
The range for each dot can be set as follows.
1 xyLv, T∆uvLv, u'v'Lv or XYZ measurement mode ..... ∆x, ∆y and ∆Lv
2 Analyzer Mode
For G-reference .............. R/G, B/G and ∆G
For R-reference ............... ∆R, B/R and G/R
3 Flicker Mode** ................... Flicker value
The range must be set independently of 1 , 2. and 3
In the case of 1 , the range set here will be used commonly by all the modes. Thus, for instance, if ∆x and
∆y are set to 2% and ∆Lv is set to 10% for xyLv mode, ∆x and ∆y will be displayed in 2% and ∆Lv in 10%
irrespective of the measurement mode (xyLv, T∆uvLv, u'v'Lv or XYZ).
In the case of 2 , the value set for ∆G (G-reference), the value set for ∆R (R-reference), the values set for R/
G and B/G (G-reference) and those set for B/R and G/R (R-reference) will be common. Thus, for instance,
if ∆G is set to 5% and both R/G and B/G are set to 3% in the case of G-reference, ∆R will be displayed in 5%
and both B/R and G/R in 3% in the case of R-reference.
In the case of 3 , the analog display range for each dot can be set for flicker value.
● Settable range ........... 0.1 to 99%
● How to Read Analog Display
When n% range is set
For xyLv, T∆ ∆ ∆ ∆ ∆ uvLv, analyzer, u'v'Lv or XYZ mode
Red
Green
–n×8%
or lower
Below ±n%
Below ±n×2%
Below ±n×4%
Below ±n×8%
• Values displayed in the analog display section
**Flicker Mode is a function which can be used only when LED Flicker Measuring ø27 Probe(CA-P32/35) or LED Flicker Measuring ø10
Probe(CA-PS32/35) is connected.
– In 0.1% step for the range from 0.1 to 9.9%
– In 1% step for the range from 10 to 99%
+
Red
+n×8%
or higher
For xyLv, T∆ ∆ ∆ ∆ ∆ uvLv, u'v'Lv or XYZ mode
x–x
∆x =
× 100 (%)
t
x
t
y–y
t
∆y =
× 100 (%)
y
t
Lv–Lv
∆Lv =
t
× 100 (%)
Lv
t
where, xt, yt, Lvt : Target color values
x, y, Lv : Measured values
69
For flicker mode**
Red
Green
Below ±n%
Below ±n×2%
Below ±n×4%
Below ±n×8%
8% or higher
Red