Konica Minolta CA-310 Instruction Manual page 114

Display color analyzer
Table of Contents

Advertisement

Item
Receptor
Measurement area
Acceptance angle
Measurement distance
Luminance
Display range
Chromaticity
Measurement range
Accuracy
(for
white)*1
Luminance
Repeatability(2 )
Measurement range
Accuracy
*1
(temperature:23°±2°,
relative humidity:
(40±10)%))
Chromaticity
Repeatability(2 )
Measurement range
Flicker
Display range
Contrast
Accuracy
method
Repeatability(2 )
Measurement range
Accuracy
Flicker JEITA
method
*3
Repeatability(2 )
xyL
V
Measurement
speed*4
Flicker Contrast
Flicker JEITA
*3
Digital
Display
Analog
LCD
SYNC mode
Object under measurement
Memory channel
Analyzer function
Interface
Multi-point Measurement
Operating temperature/humidity range
Storage temperature/humidity range
Input voltage range
Main body
Size/weight
Probe
Standard accessories
Optional accessories
*1
The chromaticity and luminance are measured under Konica Minolta's condition (standard LCD(6500K, 9300K) is used).
*2
The luminance for monochrome is measured when the reading of luminance for white is 120 cd/m
*3
Measurement of flicker (JEITA method) is supported by SDK software.
*4 Measuring probe connected to probe connector P1 only, used USB(used RS-232C Baud rate: 38400 bps)
*5
Measured by Konica Minolta's PC (CPU: 600 MHz)
*6
" " represents alternating current available at power outlets.
The specifications and drawings given here are subject to change without prior notice.
LED Flicker Measuring Ø27 Probe
Detector: Silicon photo cell
Ø27 mm
±2.5°
30±10 mm
2
0.0001 to 1000 cd/m
Displayed in 4 or 3-digit value (Can be chosen)
2
0.0050 to 1000 cd/m
2
0.0050 to 0.0999 cd/m
2
0.1000 to 9.999 cd/m
2
10.00 to 1000 cd/m
2
0.0050 to 0.0999 cd/m
2
*1
0.1000 to 0.9999 cd/m
2
1.000 to 1000 cd/m
2
0.0500 to 1000 cd/m
2
0.0500 to 4.999 cd/m
2
5.000 to 19.99 cd/m
2
20.00 to 1000 cd/m
2
120 cd/m
2
0.0500 to 0.0999 cd/m
2
0.1000 to 0.1999 cd/m
*1
2
0.2000 to 0.4999 cd/m
2
0.5000 to 1000 cd/m
2
5 cd/m
or higher
0.0 ~ 999.9 %
±1 % (Flicker frequency: 30 Hz AC/DC 10% sine wave) ±2 % (Flicker frequency: 60 Hz AC/DC 10% sine wave)
1 % (Flicker frequency: 20 to 65 Hz AC/DC 10% sine wave)
2
5 cd/m
or higher
±0.5 dB (Flicker frequency: 30 Hz AC/DC 4% (-40 dB) sine wave)
±1.0 dB (Flicker frequency: 30 Hz AC/DC 1.2% (-50 dB) sine wave)
0.1 dB (Flicker frequency: 30 Hz AC/DC 4% (-40 dB) sine wave)
0.3 dB (Flicker frequency: 30 Hz AC/DC 1.2% (-50 dB) sine wave)
2
0.0050 to 0.0999 cd/m
2
0.1000 to 1.999 cd/m
2
2.000 to 1000 cd/m
16 (16) times/sec.
0.5 (0.3) times/sec. *5
xyL
, TΔuvL
, RGB analyze, XYZ, u'v'L
V
V
ΔxΔyΔL
, R/G B/G ΔG, ΔR B/R G/R, Flicker (Contrast method) *3
V
16 characters by 2 lines (with backlight)
NTSC, PAL, EXT, UNIV, INT
Vertical synchronization frequency: 40 to 200 Hz (Luminance or chromaticity measurement), 40 to 130 Hz (Flicker measurement)
100 channels
Standard function
USB; RS-232C (38,400 bps or below)
Max. 5 points (Use 4-Probe Expansion Board CA-B15)
Temperature: 10 to 28°C; relative humidity 70% or less with no condensation
Chromaticity change ±0.002 for white, ±0.006 for monochrome from reading of Konica Minolta's standard LCD *1, 120 cd/m
with 23°C 40%
0 to 28°C: relative humidity 70% or less with no condensation 28 to 40°C: relative humidity 40% or less with no condensation
100-240V
, 50-60 Hz, 50 VA *6
340(W)×127(H)×216(D) mm / 3.58 kg
Ø49×208 mm / 530
AC power cord, PC Software for Color Analyzer CA-SDK, Standard Hood CA-H10
4-Probe Expansion Board CA-B15, LED Flicker Measuring ø27 Probe CA-P32(2 m) CA-P35(5 m), Standard Hood CA-H10,
Standard Lens Cap CA-H11, USB cable IF-A27
2
±4%±0.0015 cd/m
2
±3%±0.0010 cd/m
2
±2%±0.0010 cd/m
2
1% + 0.0010 cd/m
2
0.2% + 0.0010 cd/m
2
0.1%+0.0010 cd/m
±0.005 for white
±0.004 for white
±0.003 for white
±0.002 for white
(±0.004 for
monochrome)*2
0.010
0.004
0.002
0.001
4(3.5) times/sec.
5(4.5) times/sec.
20(17) times/sec.
, Flicker (Contrast method) *3
V
2
112
Luminance change: ±2% of reading for white ;
2
,

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents