From the Measurement Preparation section
When performing measurement using user calibration
<User Calibration> Page 51
Not required in the case of flicker measurement**.
1. Performing White Calibration Page 52
Cannot be performed with memory channel CH00.
5
8
1
4,9
2
6,7
When the optional 4-Probe Expansion Board CA-B15 is used
User calibration is performed independently for probe
no.([P1] to [P5]) for each memory channel.
1 Press the
key to display the menu selection screen.
2 Press the
key to open the PROBE selection screen.
3 Press the
key until the desired probe no. appears.
4 Press the
key to confirm the selection.
Memory channel
Probe no.
CH01 EXT
Ad P1A
[
1. Press the MODE key to select xyLv mode.
2. Press the CH
and
desired memory channel.
3. Place the measuring probe against the display
which is displaying the known white color.
4. Press the HOLD key.
The HOLD LED will light up.
CH01
P1
5. Press the
key.
CH01 x
6. Press the
key.
0000 0000
_
7. Enter the calibration values (x, y, Lv).
Ten-key (
to
,
) ... Used to enter values.
Key ... The cursor moves in the order x → y → Lv → x.
CH01 x
3300 3000 39.50
_
Cursor
CH01
8. Press the
key.
P1
" * " mark is displayed.
(The mark will be displayed for W when the value is entered.)
CH01 EXT
9. Press the
key.
[
10. Press the HOLD key. Measurement will start.
The HOLD LED will light off.
●To change the target color after user calibration:
*1 <Setting/Changing the Target Color> Page 116
●To set an ID name:
*2 <Setting an ID Name> Page 118
●To use the analog display function:
*3 <Setting the Analog Display Range> Page 118
3
]
keys to select the
U-CAL
W
R
G
B
y
Lv P1
0.00
y
Lv P1
U-CAL
W
R
G
B
Aa P1A
]
117
Not required in the case of flicker measurement**.
2. Performing Matrix Calibration Page 54
Cannot be performed with memory channel CH00.
4
9,11
1
2
7,8
6,10
When the optional 4-Probe Expansion Board CA-B15 is used
User calibration is performed independently for probe
no. ([P1] to [P5]) for each memory channel.
1 Press the
key to display the menu selection screen.
2 Press the
key to open the PROBE selection screen.
3 Press the
key until the desired probe no. appears.
4 Press the
key to confirm the selection.
Memory channel
Probe no.
CH01 EXT
Ad P1A
[
1. Press the MODE key to select xyLv mode.
2. Press the CH
and
desired memory channel.
3. Place the measuring probe against the display and
set the display so that it can display known RGBW.
CH01
4. Press the
key.
P1
5. Cause the display to show red (green), (blue),
(white).
6. Press the HOLD key.
The HOLD LED will light up.
7. Press the
(
,
,
CH01 x
0000 0000
_
8. Enter the calibration values (x, y, Lv) for R.
Ten-key (
to
,
Key ... The cursor moves in the order x → y → Lv → x.
CH01 x
3300 3000 39.50
_
9. Press the
key.
CH01
P1
(The same mark will be displayed for G, B and W
10. Press the HOLD key.
The HOLD LED will light off. Measurement will start.
* Repeat steps 4 to 10 for G, B and W.
* When "*" is displayed for R, G, B and W,
indicating that entry of all the values is complete,
CH01 EXT
11. Press the
key.
[
Matrix calibration will be performed.
To the Measurement section
3
5
]
keys to select the
U-CAL
W
R
G
B
) key.
y
Lv P1
0.00
) ... Used to enter values.
y
Lv P1
Cursor
U-CAL
W
R
G
B
" * " mark is displayed.
when their values are entered.)
Am P1A
]
Page 75