From the Measurement Preparation section
When performing measurement in analyzer mode
<Inputting the RGB Emission Characteristic for
Analyzer mode> Page 59
Can be set to all the memory channels.
7
1
2
4
6
5
When the optional 4-Probe Expansion Board CA-B15 is used
User calibration is performed independently for probe
connectors ([P1] to [P5]) for each memory channel.
1 Press the
key to display the menu selection screen.
2 Press the
key to open the PROBE selection screen.
3 Press the
key until the desired probe no. appears.
4 Press the
key to confirm the selection.
Memory channel
CH01 EXT
[
1. Press the MODE key to select analyzer mode.
2. Press the CH
and
desired memory channel.
3. Press the
key.
4. Input the emission characteristic of R.
1 Place the measuring probe against the display, which is now emitting
monochrome light of R.
2 Press the
key.
5. Input the emission characteristic of G.
1 Place the measuring probe against the display, which is now emitting
monochrome light of G.
2 Press the
key.
6. Input the emission characteristic of B.
1 Place the measuring probe against the display, which is now emitting
monochrome light of B.
2 Press the
key.
7. Input the emission characteristic of W.
1 Place the measuring probe against the display, which is now emitting
monochrome light of W.
2 Press the
key.
8. Press the
key.
The RGB emission characteristic for the display and
target color will be set.
●To change the target color after user calibration:
*1 <Setting/Changing the Target Color> Page 116
●To set an ID name:
*2 <Setting an ID Name> Page 118
●To use the analog display function:
*3 <Setting the Analog Display Range> Page 118
To the Measurement section
3
4,5,6,7
8
Probe no.
Ad P1A
]
keys to select the
CH01
ALZ
P1
W
R
G
B
In the LCD display section, the " * " mark will appear on the left of "R".
CH01
ALZ
P1
W
R
G
B
In the LCD display section, the " * " mark will appear on the left of "G".
CH01
ALZ
P1
W
R
G
B
In the LCD display section, the " * " mark will appear on the left of "B".
CH01
ALZ
P1
W
R
G
B
In the LCD display section, the " * " mark will appear on the left of "W".
CH01 EXT
Ad P1A
[
]
Page 75
118
*2
<Setting an ID Name> Page 67
Can be set to all the memory channels.
1
1.Press the CH
and
desired memory channel.
2. Press the
key.
3. Press the
key to open the ID name input
screen.
4. Enter the desired ID name.
Number-key (
to
,
Key .. Can be used to enter an alphabet,
hyphen (-) and space. Pressing this
key again will restore the original
function of the number-key.
Key .. Moves the cursor to the right each
time this key is pressed.
5. Press the
key.
*3
<Setting the Analog Display Range> Page 69
1
1.Press the MODE key to select the measurement
mode for which you want to set the range.
2. Press the
key.
3. Press the
key to open the RANGE setting
screen.
4. Enter the desired range value.
Ten-key (
to
,
Key .. Moves the cursor to the right each
time this key is pressed.
5. Press the
key.
2
3
5
4
keys to select the
Memory channel
CH01 EXT
Ad P1A
[
]
MENU : SELECT
PUSH SPACE KEY
)Used to enter values.
CH01 EXT
Ad P1
[EXT D-1.50]
_
Cursor ID name
2
3
5
4
MENU : SELECT
PUSH SPACE KEY
RANGE x,y
Lv
(%) 10
10
_
) ... Used to enter values.
For analyzer mode (G reference)
RANGE G
B/G,R/G
(%) 10
10
_
For analyzer mode (R reference)
RANGE R
B/G,R/G
(%) 10
10
_
For flicker mode**
RANGE
FMA
(%)
10
_