Konica Minolta CA-310 Instruction Manual page 84

Display color analyzer
Table of Contents

Advertisement

[Operating Procedure]
1
3
4
1
.
Set the POWER switch to ON.
2
.
Set the 0-CAL ring of the measuring probe
to the 0-CAL position.
Never direct the measuring probe toward a high-inten-
sity illuminant.
When the optional 4-Probe Expansion Board CA-B15 is used
Set the 0-CAL ring of every measuring probe to the 0-
CAL position. Zero calibration will not be performed
correctly if the 0-CAL ring of any of the measuring
probes is not set to the 0-CAL position.
3
.
Press the 0-CAL key.
After zero calibration is complete, set the 0-CAL ring
to the MEAS position and start measurement.
4
.
Press the MODE key to select analyzer measure-
ment mode (RGB).
5
.
Press the MEMORY CH
the memory channel for which the RGB emission
characteristic for analyzer mode has been set (page
59).
When the optional 4-Probe Expansion Board CA-B15 is used
Select the probe no. for which the RGB emission characteristic for the analyzer mode has been input.
1 Press the
The LCD display section will switch to the menu selection screen.
2 Press the
Each time the
order PROBE → SYNC → ID Name input → RANGE → Mea-
surement Speed → Number of Digits → Calibration Standard
→ RS232C Baud Rate → PROBE.
3 Press the
Each time the
order [P1] ....
4 Press the
* By default (factory setting), the instrument is set so that [P1] will be selected automati-
cally when the POWER switch is set to ON( | ).
6 .
Place the measuring probe against the display and
take measurement.
5
and
key.
key to open the PROBE selection screen.
key is pressed, the screen will switch in the
key to display the probe no. you want to select.
key is pressed, the probe no. switches in the
key to confirm the selection.
7
5
2,3
6
DARKEN PROBE
PUSH 0-CAL KEY
ZERO CALIBRATION
CH00 EXT
[
keys to select
Memory channel
Menu selection screen
PROBE selection screen
82
Block entry of light
Place the probe against the display
Message displayed
when the POWER
switch is set to ON
Press the 0-CAL key.
During zero calibration
Ad P1A
]
End of zero calibration
Probe no.
CH01
EXT
P3A
[
]
MENU : SELECT
PUSH SPACE KEY
SELECT : PROBE
P1 35881112 A
SELECT : PROBE
P3 35881113
Probe no.
Press the
key until the
desired probe
no. appears.