The second calibration that is performed is the automatic blood dimness
calibration. When its finished, the following calibration complete screen
will be displayed:
Press the [CONFIRM] key to save the data. The screen will change.
5.
Press the [CONFIRM] key again to complete the calibration and return to
the Calibrate Sensors screen.
Page 48
Fresenius 2008K Calibration Procedures
P/N 507296 Rev. A