Fresenius Medical Care 2008k Technical Manual page 76

Hemodialysis machine
Hide thumbs Also See for 2008k:
Table of Contents

Advertisement

4.
Push the syringe plunger in to show a pressure of 200mmHg on the external
pressure gauge.
5.
While holding this pressure, do the following:
• Use the [
Pressure data box on the 2008K display to indicate 200mmHg.
• Press the [CONFIRM] key and the 2008K display will change.
• Press the [Start/Stop] key on the blood pump and its display will
change back to the alternating display:
• Verify that the Arterial Pressure and Pressure at Actuator Board
meter boxes are within 10 mmHg of each other.
N
Note: After pressing the [CONFIRM] key, a delay of approximately
8 seconds may occur before the Pressure at Actuator Board data box
changes to be within 10 mmHg of the Arterial Pressure data box on the
display screen.
Pressurize the P
6.
Verify that the Arterial Pressure and Pressure at Actuator Board meter
7.
boxes can reach at least 290mmHg (510mmHg for postpump) and that they are
within 10mmHg of each other.
Verify that the Arterial Pressure and Pressure at Actuator Board meter
boxes can reach at least -310mmHg (-90mmHg for postpump) and that they are
within 10mmHg of each other.
Press the [CONFIRM] key to save the data. The screen will change.
8.
U
Caution: The following step must be performed on the blood pump or the
calibration data you just collected and tested will be lost when the machine
is turned off.
Fresenius 2008K Calibration Procedures
=
>
] keys on the blood pump to adjust the Arterial
] and [
port by either pushing or pulling on the syringe.
ART.
P/N 507296 Rev. A
Page 29