Fresenius Medical Care 2008k Technical Manual page 119

Hemodialysis machine
Hide thumbs Also See for 2008k:
Table of Contents

Advertisement

OFF AFTER HEAT DISIN: This option is used to select if the machine
Line S:
will power off after a heat disinfect program. This setting will toggle from
Yes to No by pressing the Off after Heat Disin button. If Yes is selected,
the machine will power off after a heat disinfect program. If No is selected,
the machine will NOT power off after a heat disinfect program. Once the
option is set press the [CONFIRM] key and the X will change from gray to
blue. Select Yes or No as dictated by facility management. Record this
setting on Installation Checklist.
FDS08 BEACON: This option is used to enable or disable the
Line T:
communication between the FDS08 system and the Beacon. This setting will
toggle from Yes to No by pressing the FDS08 Beacon button. If Yes is
selected, the Beacon lights will be controlled by the FDS08 system. If No is
selected, the Beacon lights will be controlled by the color of the 2008K status
box. Once the option is set press the [CONFIRM] key and the X will change
from gray to blue. Select Yes or No as dictated by facility management.
Record this setting on Installation Checklist.
Line U:
USE CMS208K: This option is used to enable or disable the
communication with the CMS208K system. This setting will toggle from
Yes to No by pressing the Use CMS208K button. Once the option is set
press the [CONFIRM] key and the X will change from gray to blue. Select
Yes or No as dictated by facility management. Record this setting on
Installation Checklist.
BVM: This option is used to enable or disable the use of the blood volume
Line V:
monitor option. This setting will toggle from Yes to No by pressing the
BVM button. Once the option is set press the [CONFIRM] key and the X
will change from gray to blue. Select Yes or No as dictated by facility
management. Record this setting on Installation Checklist.
Line W:
CLEARANCE: This option is used to enable or disable the online
clearance option. This setting will toggle from Yes to No by pressing the
Clearance button. Once the option is set press the [CONFIRM] key and
the X will change from gray to blue. Select Yes or No as dictated by
facility management. Record this setting on Installation Checklist.
® ®
DIASAFE
Line X:
®
DIASAFE
pressing the Diasafe Filter button. Once the option is set press the
[CONFIRM] key and the X will change from gray to blue. Select Yes or
No as dictated by facility management. Record this setting on Installation
Checklist.
FILTER: This option is used to enable or disable the
filter option. This setting will toggle from Yes to No by