12. Move the carriage all the way up to the end of its travel.
13. Select the RATE data button. Use the [
9.9 ml/hr then press the [CONFIRM] key.
14. Press the [Heparin on/off] key to start the pump. The green lamp above
the key should light.
15. Verify that a HEPARIN PUMP ALARM is displayed on the status line
within 2 minutes.
N
Note: If the machine is not alarm free, another message of higher
priority may be displayed on the status line.
3.17

BLOOD PUMP

Clean and test the arterial blood pump module as follows:
1. Remove the blood pump rotor by opening the door, pulling out the
handle and turning the rotor 90 degrees
2. Clean the rollers with a cloth dampened only with water.
3. Clean the inside of the blood pump housing with the damp cloth.
Compressed air may be used if it is available.
4. Remove the blood pump module from the machine and inspect the
motor gearbox. In heavy use, some oil may accumulate on the gearbox
housing. Wipe the housing clean.
5. Reassemble the blood pump module and replace it in the machine.
6. Place the machine in Dialysis Mode and clear all blood alarms.
7. Start the blood pump if it is not already running.
Page 32
Fresenius 2008K Preventive Maintenance Procedures
P/N 507297 Rev. A
=
>