2008K Technician's Manual
SPECIFICATIONS
2008K HEMODIALYSIS MACHINE
Approximately 54 cm wide by 63 cm deep
133 cm
Approximately 73 kg
Single phase AC 120 V ±10% 60 Hz ±3 Hz must be
connected to a circuit which is equipped with a hospital
grade receptacle and is protected by circuit breaker and
ground fault circuit interrupter (GFCI). Resistance from
chassis to ground must be < 0.2 ohm.
Does not exceed 12.5 amps
6.3 amp medium blow fuse, 2 each
16 amp double pole rocker switch circuit breaker for
heater
A.
Alarm output normally open and normally closed
switch contacts reflecting alarm state. 24 V, 0.2
amp max.
B.
Alarm input
C.
External alarm light or traffic light beacon
Isolated RS232 and leakage current isolation per UL 544
between the machine and external computer
600 to 700 BTU/hr.
Section I – Specifications
I-1