Concentrate Supply (cont.)
Adjustment Range
Monitoring Conductivity
Range of Display
Dialysate Heating
Nominal Value of Temperature
Temperature Display
Heat Disinfection
Temperature
Flow Rate
Time
Chemical Disinfection
Temperature
Flow Rate
Time
Blood Pump
Display of flow rate
Internal diameter of pump segment
Tube length
Level adjust
Power outage use
2008K Technician's Manual
20 to 40 mEq/L Bicarbonate
Temperature compensated conductivity display with
automatically set alarm windows ±0.5 mS/cm around
calculated conductivity. User can adjust an additional
±0.5mS/cm within this range. Conductivity is based on
the concentrates' compositional data entered in the
Dialysate screen at the standard temperature of 25°C.
10.0 to 17.0 mS/cm corrected to 25° C. Alarm limits will
not go below 13.0 or above 16.0 mS/cm.
35 to 39° C, ( selectable in 0.1 ° C steps )
Range 35 to 39° C with alarm limit window
automatically adjusted to 2° C above and below set point.
Alarm window will not go below 30° or above 41° C.
Heater 1.3 kW, electronically controlled.
83 ± 8° C at NTC 3
600 ml/min.
Pre-rinse either 7min. @ 600ml/min or 20min. @
300ml/min. (user selectable)
Between 10 and 60 minutes (internally selectable)
37° C (set point applicable)
620 ml/min.
Between 10 and 60 minutes (internally selectable)
20 to 600 ml/min, depending on pump segment ID.
2 to 10 mm (0.08" to 0.4")
32 cm minimum (12-5/8")
Up only
The pump can be manually operated with a hand crank.
Section I – Specifications
I-3